Setting Up Measurement Conditions and Display
Setting Up Beep
Setting Up Beep
Setting up conditions to make a beep sound
To turn ON/OFF the beep sound, use one of the following commands. You can
use either of these two commands since both functions are identical.
•
•
To generate the beep sound regardless of ON/OFF setting, use
:SYSTem:BEEPer[:IMMediate]
sound from five different types of available tones, use :SYSTem:BEEPer:TONE
on page
To set the conditions to make a beep sound according to the comparator
sorting result, use the following command.
•
The table below shows the relationship between the settings made by the
above commands and the condition to make a beep.
:CALCulate1:CO
MParator:BEEPe
:SYSTem:BEEPer
Making a beep sound
To make a beep sound, use the following command:
•
46
:CALCulate1:COMParator:BEEPer[:STATe]
:SYSTem:BEEPer:STATe
274.
:CALCulate1:COMParator:BEEPer:CONDition
Setting of
:CALCulate1:C
OMParator:BEE
or
r[:STATe]
Per:CONDition
:STATe
OFF
ON
:SYSTem:BEEPer[:IMMediate]
on page 273
on page
273. To change the tone of beep
Setting of
Cond ition to make a beep
———
Never make a beep sound
When wrong
•
FAIL
key operation is
performed.
When an error,
•
alarm, or other
message is
output.
PASS
on page 273
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
on page 174
on page 173
When the sorting
•
judgment result of
the comparator is
OUT_OF_BIN, OVLD,
LOWC_OR_NC or
AUX_BIN.
When the sorting
•
judgment result of
the comparator is
within BIN1 to BIN9.