Command Reference
E4981A SCPI Commands

:TEST:SCANner:EOM

Syntax
:TEST:SCANner:EOM {ON|OFF|1|0}
Description
This command sets the Scanner End of Measurement (EOM) signal for test purpose
and can be used to check/troubleshoot the Test pin signal of Scanner interface. (No
Query)
When
can be controlled/read by: TEST:SCAN:xxxx commands. After completing
the test,
pin signal is generated according to the actual measurement result.
Parameters
ON
OFF
Related commands
:TEST:SCANner:CH?
:TEST:SCANner:INDex
:TEST:SCANner:MODE
:TEST:SCANner:TRIGger?
:TEST:SCANner:VALid?
Equivalent key
No equivalent key is available on the front panel.
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
:TEST:SCANner:MODE
:TEST:SCANner:MODE
Description
Sets Scanner EOM status to ON.
Sets Scanner EOM status to OFF.
on page 298
on page 300
on page 301
on page 302
on page 302
is set to ON, the interface pin signal value
should set to OFF so that scanner interface
299