Test:scanner:index - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

:TEST:SCANner:INDex

Syntax
:TEST:SCANner:INDex {ON|OFF|1|0}
Description
This command sets the Scanner Index signal for test purpose and can be used
to check/troubleshoot the Test pin signal of Scanner interface. (No Query)
When
can be controlled/read by: TEST:SCAN:xxxx commands. After completing
the test,
pin signal is generated according to the actual measurement result.
Parameters
ON
OFF
Related commands
:TEST:SCANner:CH?
:TEST:SCANner:EOM
:TEST:SCANner:MODE
:TEST:SCANner:TRIGger?
:TEST:SCANner:VALid?
Equivalent key
No equivalent key is available on the front panel.
sequence
300
:TEST:SCANner:MODE
:TEST:SCANner:MODE
Description
Sets Scanner Index signal status to ON.
Sets Scanner Index status to OFF.
on page 298
on page 299
on page 301
on page 302
on page 302
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
is set to ON, the interface pin signal value
should set to OFF so that scanner interface