Command Reference
E4981A SCPI Commands
:TEST:HANDler:STATus:OVLD
Syntax
:TEST:HANDler:STATus:OVLD {ON|OFF|1|0}
Description
This command sets the Handler Overload signal status for test purpose and
can be used to check/troubleshoot the Test pin signal of Handler interface. (No
Query)
When
can be controlled/read by: TEST:HAND:xxxx commands. After completing
the test,
pin signal is generated according to the actual measurement result.
Parameters
ON
OFF
Related commands
:TEST:HANDler:STATus:ALARm
:TEST:HANDler:STATus:INDex
:TEST:HANDler:STATus:NC
:TEST:HANDler:STATus:RDYTrig
Equivalent key
No equivalent key is available on the front panel.
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
:TEST:HANDler:MODE
:TEST:HANDler:MODE
Description
Sets Handler Overload signal status to ON.
Sets Handler Overload signal status to OFF.
is set to ON, the interface pin signal value
should set to OFF so that handler interface
on page 291
on page 293
on page 294
on page 296
295