[:Sense][:Fimpedance]:Aperture:time; [:Sense][:Fimpedance]:Contact1:Verify:buf1:Data - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe][:FIMPedance]:APERture:TIME

Syntax
[:SENSe][:FIMPedance]:APERture:TIME {1|2|4|6|8}
[:SENSe][:FIMPedance]:APERture:TIME?
Description
Specifies the measurement time.
For information on the specific measurement time of each mode, see
"Specification and Supplemental Performance Characteristics" in the
Operation Manual.
Parameters
1, 2, 4, 6, 8
Query response
{1|2|4|6|8}<newline><^END>
Equivalent key
[Meas Setup] - MEAS TIME - INCR+|DECR-
sequence

[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:DATA?

Syntax
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:DATA?
Description
Gets the data from buffer1 Contact1. Maximum buffer size is 1000. (Query
Only)
The transfer format is based on the setting of :FORMat[:DATA].
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:FEED:CONTrol:INTer
val
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:FEED:CONTrol[:STAT
e]
[:SENSe][:FIMPedance]:CONTact1:VERify:BUF1:POINts
Equivalent key
No equivalent key is available on the front panel.
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Description
Specifies measurement speed (time).
on page 254
on page 255
on page 256
253

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4981a/001E4981a/002

Table of Contents