Calculate1:Comparator:count[:State] - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

:CALCulate1:COMParator:COUNt[:STATe]

Syntax
:CALCulate1:COMParator:COUNt[:STATe] {ON|OFF|1|0}
:CALCulate1:COMParator:COUNt[:STATe]?
Description
Turns ON/OFF the BIN counter function of the comparator function.
If you turn on this function, the number of DUTs sorted into each BIN based on
the comparator sorting result is counted. The maximum count is 999999. If this
is exceeded, the count value will not be updated but remain at 999999.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands
:CALCulate1:COMParator:COUNt:CLEar
:CALCulate1:COMParator:COUNt:DATA?
:CALCulate1:COMParator:COUNt:OVLD?
:CALCulate1:COMParator:COUNt:MULTiple:DATA?
:CALCulate1:COMParator:COUNt:MULTiple:OVLD?
Equivalent key
[Display Format] - BIN COUNT - COUNT - COUNT ON|COUNT OFF
sequence
180
Description
Turns ON the BIN counter function.
Turns OFF the BIN counter function.
on page 175
on page 176
on page 179
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
on page 177
on page 179