[:Sense]:Correction:collect:stan3:Range:auto - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe]:CORRection:COLLect:STAN3:RANGe:AUTO

Syntax
[:SENSe]:CORRection:COLLect:STAN3:RANGe:AUTO {ON|OFF|1|0}
[:SENSe]:CORRection:COLLect:STAN3:RANGe:AUTO?
Description
Sets on/off of auto ranging when the load correction is executed. When this
function is set at off, the range which is selected by
[:SENSe][:FIMPedance]:RANGe[:UPPer]
For auto ranging when a measurement is executed, use
[:SENSe][:FIMPedance]:RANGe:AUTO
The functionality of this command is same as
[:SENSe]:CORRection:COLLect:LOAD:RANGe:AUTO.
Parameter
ON or 1 (initial
setting)
OFF or 0
Query response
{1|0}<newline><^END>
Related Commands
[:SENSe]:CORRection:COLLect[:ACQuire]
[:SENSe][:FIMPedance]:RANGe[:UPPer]
:SYSTem:PRESet
Equivalent key
[Meas Setup] - CORRECTION - LOAD RNG- AUTO/FIX.
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Description
Measures the load standard with auto range.
Measures the load standard measurement range which has been
set just before the load standard measurement is performed (auto
range function is invalid).
on page 284
is used for the load correction.
on page 237
on page 266
239