System:communicate:gpib[:Self]:Address; System:communicate:lan[:Self]:Address - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess

Syntax
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <numeric>
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess?
Description
Sets the GPIB address.
Parameter
Range
Initial (Factory set) value
Resolution
The value of this command is not changed with
and is only changed with Factory Default reset (available only through
Front Panel).
{numeric}<newline><^END>
Query response
Equivalent key
[System] - SYSTEM CONFIG - GPIB ADDR

:SYSTem:COMMunicate:LAN[:SELF]:ADDRess

Syntax
:SYSTem:COMMunicate:LAN[:SELF]:ADDRess <String>
:SYSTem:COMMunicate:LAN[:SELF]:ADDRess?
Description
Sets the static IP address.
Parameter
Initial (Factory set) value
The value of this command is not changed with
and is only changed with Factory Default reset (available only through
Front Panel).
{"string"}<newline><^END>
Query response
Equivalent key
[System] - SYSTEM CONFIG - MANUAL IP ADDR - ENTER
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
<Numeric>
0 to 30
17
1
*RST
<String>
"192.168.1.101"
*RST
and
:SYSTem:PRESet
and
:SYSTem:PRESet