Turning On/Off Aux Bin Function; Setting Up Condition To Make A Beep Sound - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Sorting Based on Measured Result (Comparator Function)
Setting Up Comparator Function

Turning ON/OFF AUX BIN function

To turn ON/OFF the AUX BIN function, use the following command.
Depending on the ON/OFF state of the AUX BIN function, the sorting result
when the value of the secondary parameter exceeds the limit range varies as
shown in
Table 7-1
Sorting result when measured secondary parameter value exceeds limit
range

Setting up condition to make a beep sound

You can select the condition that produces a beep sound based on the
comparator sorting judgment result from the following:
To set the condition that makes a beep sound, use the following command.
You can also disable the beep sound. To turn ON/OFF the beep, use either of
the following commands (they have the same function).
76
:CALCulate1:COMParator:AUXBin
Table
7-1.
Primary parameter sorting resul t
One of BIN1 to BIN9
Not sorted to any BIN
Beep when the sorting judgment result is OUT_OF_BINS or AUX_BIN
Beep when the sorting judgment result is BIN1 to BIN9
:CALCulate1:COMParator:BEEPer:CONDition
:CALCulate1:COMParator:BEEPer[:STATe]
:SYSTem:BEEPer:STATe
on page 172
AUX BIN function
OFF
ON
Not applicable
on page 273
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Sorting resul t
OUT_OF_BINS
AUX_BIN
OUT_OF_BINS
on page 173
on page 174