[:Sense]:Correction:load[:State] - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands
Equivalent key
[Meas Setup] - CORRECTION - OPEN - A|B
sequence
[Meas Setup] - CORRECTION - SHORT- A|B
[Meas Setup] - CORRECTION - LOAD- A|B

[:SENSe]:CORRection:LOAD[:STATe]

Syntax
[:SENSe]:CORRection:LOAD[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:LOAD[:STATe]?
Description
Turns ON/OFF the LOAD correction function.
With the LOAD correction set to ON, if you change the cable length (set with
the
the
changed to OFF.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands
:CALibration:CABLe[:LENGth]
:SYSTem:FSHift
[:SENSe]:CORRection:COLLect[:ACQuire]
Equivalent key
[Meas Setup] - CORRECTION - LOAD - ON/OFF
sequence
242
:CALibration:CABLe[:LENGth]
command), the LOAD correction is automatically
:SYSTem:FSHift
on page 281
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
command) or frequency shift (set with
Description
Turns ON the LOAD correction.
Turns OFF the LOAD correction.
on page 171
on page 237