[:Sense][:Fimpedance]:Creject:limit - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe][:FIMPedance]:CREJect:LIMit

Syntax
[:SENSe][:FIMPedance]:CREJect:LIMit <numeric>
[:SENSe][:FIMPedance]:CREJect:LIMit
Description
Sets the boundary value (percentage of the measurement range) within the
detection range of Low C when you turn on the Low C reject function. The
measurement range to which the set value is applied differs depending on the
setup of the measurement range mode as shown below:
For example, if you make a measurement with the range fixed to the 1 µF range
and specify 1%, Low C is detected if the measured value of the primary
parameter (Cs or Cp) is 10 nF or less.
For information on the screen display, GPIB/USB/LAN output, and so on when
Low C is detected, refer to Appendix D, "Operations when Overload, No
Contact, or Low C is Detected,".
Parameters
Description
Range
Initial value
Unit
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
When in auto-range mode (ON has been specified with the
[:SENSe][:FIMPedance]:RANGe:AUTO
When the measurement frequency is in the 120 Hz: 10E-9 F (100 pF) range
When the measurement frequency is in the 1 kHz: 100E-12 F (100 pF) range
When the measurement frequency is in the 1 MHz: 1E-12 F (1 pF) range
When in the fixed range mode (OFF has been specified with the
[:SENSe][:FIMPedance]:RANGe:AUTO
<numeric>
Boundary value
0 to 10
0
% (percent)
command):
command):
263