Command Reference
E4981A SCPI Commands

:CALCulate1:COMParator:SECondary:LIMit

Syntax
:CALCulate1:COMParator:SECondary:LIMit <numeric 1>,<numeric 2>
:CALCulate1:COMParator:SECondary:LIMit?
Description
Sets the limit range for the secondary parameter used in the comparator
function.
This command only sets the limit range. To activate the set limit range, use the
:CALCulate1:COMParator:SECondary:STATe
judgment for the measurement result of the secondary parameter.
Parameters
Description
Range
Initial value
Unit
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands
:CALCulate1:COMParator:SECondary:STATe
:CALCulate1:COMParator:PRIMary:BIN{1-9}[:LIMIT]
:CALCulate1:COMParator[:STATe]
Equivalent key
[Meas Setup] - LIMIT TABLE - BIN|2nd - LOW, HIGH
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
<numeric 1>
The lower limit value of the limit
range.
-99.9999E9 to 99.9999E9
0
Depends on the type of
parameter.
on page 187
command to enable sorting
<numeric 2>
The upper limit value of the
limit range.
-99.9999E9 to 99.9999E9
0
Depends on the type of
parameter.
on page 186
on page 182