Command Reference
E4981A SCPI Commands
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe]
Syntax
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATE] {ON|OFF|1|0}
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe]?
Description
Determines whether to use the contact check function.
Contact check function is not available at 1 MHz measurement frequency.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2
Equivalent key
[Meas Setup] - CONT CHK1- ON|OFF
sequence
260
Description
Enables the contact check function.
Disables the contact check function.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
on page 261
on page 262