[:Sense]:Correction:ckit:stan3:Format - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe]:CORRection:CKIT:STAN3:FORMat

Syntax
[:SENSe]:CORRection:CKIT:STAN3:FORMat
[:SENSe]:CORRection:CKIT:STAN3:FORMat?
Description
Sets the types of parameters used to define the standard for LOAD correction.
Parameters
CPD
(initial value)
CPQ
CPG
CPRP
CSD
CSQ
CSRS
For details on Cp and Cs, refer to :CALCulate1:FORMat
details on D, Q, Rp, and Rs, refer to :CALCulate2:FORMat
Query response
{CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS}<newline><^END>
Related commands
[:SENSe]:CORRection:CKIT:STAN3[:DATA]
Equivalent key
[Meas Setup] - CORRECTION - LOAD- Cp
sequence
236
Description
Specifies Cp as the primary parameter, D as the secondary parameter.
Specifies Cp as the primary parameter, Q as the secondary parameter.
Specifies Cp as the primary parameter, G as the secondary parameter.
Specifies Cp as the primary parameter, Rp as the secondary parameter.
Specifies Cs as the primary parameter, D as the secondary parameter.
Specifies Cs as the primary parameter, Q as the secondary parameter.
Specifies Cs as the primary parameter, Rs as the secondary parameter.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
{CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS}
on page
on page
on page 235
{CPD|CPQ|CPG|CPRP}/ Cs{CSD|CSQ|CSRS}
188. For
191.

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4981a/001E4981a/002

Table of Contents