Command Reference
E4981A SCPI Commands
:SYSTem:BEEPer[:IMMediate]
Syntax
:SYSTem:BEEPer[:IMMediate]
Description
Produces a beep sound.
If the beep sound is disabled (OFF has been specified with the
:SYSTem:BEEPer:STATe
execute this command. (No query)
Related commands
:SYSTem:BEEPer:STATe
Equivalent key
No equivalent key is available on the front panel.
sequence
:SYSTem:BEEPer:STATe
Syntax
:SYSTem:BEEPer:STATe {ON|OFF|1|0}
:SYSTem:BEEPer:STATe?
Description
Turns ON/OFF the beep output.
This command has the same function as the
:CALCulate1:COMParator:BEEPer[:STATe]
Parameters
ON or 1 (initial value)
OFF or 0
Query response
{1|0}<newline><^END>
Related commands
:CALCulate1:COMParator:BEEPer[:STATe]
Equivalent key
[System] - SYSTEM CONFIG - BEEPER ENABLED - ON|OFF
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
command), no beep sound is produced even if you
on page 273
command.
Description
Enables the beep sound.
Disables the beep sound.
on page 174
273