[:Sense]:Correction:multiple:channel - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe]:CORRection:MULTiple:CHANnel

Syntax
[:SENSe]:CORRection:MULTiple:CHANnel <numeric>
[:SENSe]:CORRection:MULTiple:CHANnel?
Description
Specifies a channel number used in the multi-correction function.
You can also specify the channel number via the scanner interface.
A channel number specified through the interface overrides a channel
number selected with this command.
Parameters
Description
Range
Initial value
Resolution
If the specified parameter is out of the allowable setup range, an error occurs.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Equivalent key
[Meas Setup] - CORRECTION - CH
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
<numeric>
The desired channel number.
0 to 255
0
1