Keysight E4416A Installation Manual

Epm-p series, peak and average power meters
Hide thumbs Also See for E4416A:

Advertisement

Quick Links

Keysight E4416A and
E4417A EPM-P Series Peak
and Average Power Meters
Installation
Guide

Advertisement

Table of Contents
loading

Summary of Contents for Keysight E4416A

  • Page 1 Keysight E4416A and E4417A EPM-P Series Peak and Average Power Meters Installation Guide...
  • Page 2: Safety Information

    Conformity. understood and met. where in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With WARNING respect to any technical data as defined by FAR 2.101, pursuant to FAR...
  • Page 3: Cooling, Cleaning, And Usage Requirements

    This instrument must be used in a normal condition only (in which all means for protection are intact). This instrument is designed for use in Installation Category II and Pollution CAUTION Degree 2 per IEC61010 and 60664 respectively. Keysight E4416A and E4417A Installation Guide...
  • Page 4: Safety Symbols

    NOT isolated from the mains when the switch is pressed. On (mains supply) To isolate the instrument, the mains coupler (mains input cord) should be removed from the power supply. Protective earth (ground) terminal Keysight E4416A and E4417A Installation Guide...
  • Page 5: Safety Considerations

    Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements.
  • Page 6 DO NOT substitute parts or modify equipment: Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the product. Return the product to a Keysight Sales and Service Office for service and repair to ensure the safety features are maintained.
  • Page 7: Compliance And Markings

    Control and Laboratory Use, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition. Keysight E4416A and E4417A Installation Guide...
  • Page 8: Regulatory Information

    This statement is provided to comply with the requirements of the German Sound DIN 45635 T. 19 (Typprufung). – Sound Pressure LpA < 70 dB. – At operator position. – Normal operation. – According to ISO 7779 (Type Test). Keysight E4416A and E4417A Installation Guide...
  • Page 9: Responsibilities Of The Customer

    – Access to and use of all information and facilities determined necessary by Keysight to service and/or maintain the products. (Insofar as these items may contain proprietary or classified information, the customer shall assume full responsibility for safeguarding and protection from wrongful use.) –...
  • Page 10: Regulatory Markings

    For regulatory markings information, refer to the EPM-P Series Peak and Average Power Meters User’s Guide. Sales and Technical Support To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites: – www.keysight.com/find/epm...
  • Page 11: Table Of Contents

    ......... .32 Keysight E4416A and E4417A Installation Guide...
  • Page 12 ............40 Keysight E4416A and E4417A Installation Guide...
  • Page 13 Keysight E4416A and E4417A EPM-P Series Peak and Average Power Meters Installation Guide Getting Started Welcome Documentation Information Power Meter and Sensor Capability Adjusting the Carrying Handle Turning the Power Meter On Front Panel Keys and Connections Connecting a Power Sensor...
  • Page 14: Getting Started

    If there is mechanical damage, notify the nearest Keysight Technologies office. Keep the damaged shipping materials (if any) for inspection by the carrier and a Keysight representative. If required, you can refer to “Sales and Technical Support”...
  • Page 15: Documentation Information

    This guide is only part of the information supplied. The documentation consists of: – The installation guide (this book) - Shows you how to check your power meter, switch it on and connect it to a Keysight power sensor. This information is presented in English and Japanese languages.
  • Page 16: Conventions Used In This Guide

    This is used to represent a displayed message. Message This is used to represent a parameter, value, or title. Parameter Keysight E4416A and E4417A Installation Guide...
  • Page 17: Power Meter And Sensor Capability

    Getting Started Power Meter and Sensor Capability Your E4416A or E4417A power meter is compatible with Keysight E-Series E9320, E-Series E9300, E-Series E4400, 8480 Series and N8480 Series power sensors. However, not all sensor and meter combinations have the same features or capabilities.
  • Page 18: Adjusting The Carrying Handle

    Pull the handle outwards, rotate it to the required position and release it into one of the three locks. If you want to remove the handle, refer to “Rack Mounting the Power Meter” page 33. Keysight E4416A and E4417A Installation Guide...
  • Page 19: Turning The Power Meter On

    The instrument has an autoranging power supply. Ensure the supply voltage is CAUTION within the range 85 to 264 V and 47 - 440Hz. 1 Connect the power cord. 2 Check the orange LED is lit. POWER Keysight E4416A and E4417A Installation Guide...
  • Page 20 Getting Started 3 Turn the meter on and confirm the green LED is lit. POWER 4 The meter automatically steps through a self test routine..and is now ready for use. Keysight E4416A and E4417A Installation Guide...
  • Page 21: What Can Go Wrong

    If there are any self-test failures the meter is defective. Contact your Fails sel f test(s) nearest Keysight Service Center (Refer to “Sales and Technical Support” on page 10). 5 Line input fuse and spare. In-line fuse Spare fuse Keysight E4416A and E4417A Installation Guide...
  • Page 22: Front Panel Keys And Connections

    It also enables you to take control of the meter from the front panel when operating via the remote interfaces (when Local Lock Out is not enabled). Keysight E4416A and E4417A Installation Guide...
  • Page 23 E-Series E9320A sensor is connected, all the menu keys are disabled (grayed out). Use this key to configure the selected measurement. Meas Setup Meas Meas Use this key together with to configure measurement Display Setup displays. Keysight E4416A and E4417A Installation Guide...
  • Page 24 The arrow keys are used to select measurement displays and to select and change parameters such as instrument state names and offset values. The user’s guide shows how these keys are used in more detail. Keysight E4416A and E4417A Installation Guide...
  • Page 25 The sensor input connectors (Keysight E4417A shown, the CHANNEL E4416A has one input). If the meter is configured with Option 002 or Option 003, the connectors are fitted to the rear panel. Keysight E4416A and E4417A Installation Guide...
  • Page 26: Connecting A Power Sensor

    Connecting a Power Sensor Using the correct cable, any Keysight diode based power sensor can be connected to the E4416A or E4417A power meters. (The E9288 Series cables are color coded to help distinguish them from the 11730 Series.) Power Sensor...
  • Page 27 Read ing Sensor Message does not appear when connecting an 8480 Series power sensor. 4 Confirm the display has changed to a measurement reading. Figure shows display with an E-Series E9320 power sensor connected. Keysight E4416A and E4417A Installation Guide...
  • Page 28 Zero + Cal A Zero + Cal B pop-up is displayed during the zeroing process, the Calibrating Zeroing pop-up during calibration. The meter-sensor measurement path is calibrated when the pop-up disappears. Calibrating Keysight E4416A and E4417A Installation Guide...
  • Page 29: Rear Panel Connections

    The following connections are available on the rear panel. To setup the remote interfaces, refer to “Remote Interface Configurations” on page 31. 1 GPIB 2 RS232 and RS422 RS232 RS422 1 2 3 4 5 6 7 8 9 CTS- RTS+ CTS+ RTS- Keysight E4416A and E4417A Installation Guide...
  • Page 30 TTL Output 1 TTL Output 2 TTL Input 1 TTL Input 2 Ground Ground 4 Recorder output (two outputs are fitted to dual channel meters) and Trigger Input and Output connections are made via BNC connectors. Keysight E4416A and E4417A Installation Guide...
  • Page 31: Remote Interface Configurations

    Press Enter confirm and save the new address. Pressing restores the previous Cancel value. System To ensure the GPIB interface is selected press Remote Interface Select Interface GPIB Keysight E4416A and E4417A Installation Guide...
  • Page 32: Rs232/Rs422

    4 Press to access the second page of the settings menu. More To configure the power meter to use the serial RS232 or RS422 interfaces press System , or Remote Interface Select Interface RS232 RS422 Keysight E4416A and E4417A Installation Guide...
  • Page 33: Rack Mounting The Power Meter

    Getting Started Rack Mounting the Power Meter Rack mounting one meter (Using the Option 908 rack mount kit) 1 Remove the carrying handle. 2 Remove the front and rear rubber bumpers. Keysight E4416A and E4417A Installation Guide...
  • Page 34 Getting Started 3 Fit the rack mount flanges. (Parts available separately: 5063-9240). Ready for installation. Keysight E4416A and E4417A Installation Guide...
  • Page 35 Getting Started Rack mounting two meters together (Using the Option 909 rack mount kit) 1 Remove the carrying handle. 2 Remove the front and rear rubber bumpers. 3 Hardware required to link meters together. Keysight E4416A and E4417A Installation Guide...
  • Page 36 4 Fit one small rack mounting flange to opposite sides of each power meter. (Parts available separately: 5061-9694 and 5063-9212). 5 Fit two front linking plates to each power meter. 6 Engage the linking plates at the front of the power meters. Keysight E4416A and E4417A Installation Guide...
  • Page 37 Getting Started 7 Attach the rear linking brackets. Ready for installation. Keysight E4416A and E4417A Installation Guide...
  • Page 38 Getting Started THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight E4416A and E4417A Installation Guide...
  • Page 39: Specifications

    Keysight E4416A and E4417A EPM-P Series Peak and Average Power Meters Installation Guide Characteristics and Specifications For the characteristics and specifications of the E4416A and E4417A EPM-P Series Peak and Average Power Meters, refer to the datasheet at http://literature.cdn.keysight.com/litweb/pdf/5980-1469E.pdf.
  • Page 40 Characteristics and Specifications THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight E4416A and E4417A Installation Guide...
  • Page 41 This information is subject to change without notice. Always refer to the English version at the Keysight website for the latest revision. © Keysight Technologies 2000-2017 Edition 3, November 1, 2017 Printed in Malaysia *E4416-90002* E4416-90002 www.keysight.com...

This manual is also suitable for:

E4417a

Table of Contents