Rcl; Rst - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
IEEE Common Commands

*RCL

Syntax
*RCL <numeric>
Description
Recalls the instrument setups saved in the register of the specified number on
Flash (internal) memory or USB pen drive (external memory). For details of
recalled instrument setups, refer to Operation Manual, Appendix C "Initial
Setting".(No query)
Parameters
Description
Range
Resolution
If the specified parameter is out of the allowable setup range, an error occurs.
Related commands
*SAV
Equivalent key
[Save/Recall] - No {1-9} - RECALL
sequence

*RST

Syntax
*RST
Description
Resets the instrument to the preset state.
The preset state is different from that when resetting is done using the
:SYSTem:PRESet
C "Initial Setting" (No query)
Related commands
:SYSTem:PRESet
:INITiate:CONTinuous
Equivalent key
No equivalent key is available on the front panel.
sequence
160
<numeric>
Specified number
0 to 9 (Internal Flash Memory)
10 to 19 (External USB Memory)
1
on page 161
command. For details, refer to Operation Manual, Appendix
on page 284
on page 227
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4981a/001E4981a/002

Table of Contents