Test:scanner:ch - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

:TEST:SCANner:CH?

Syntax
:TEST:SCANner:CH?
Description
This command gets the Scanner channel number for test purpose and can be
used to check/troubleshoot the Test pin signal of Scanner interface. (Query
Only)
When
can be controlled/read by: TEST:SCAN:xxxx commands. After completing
the test,
pin signal is generated according to the actual measurement result.
<numeric><newline><^END>
Query response
Related commands
:TEST:SCANner:EOM
:TEST:SCANner:INDex
:TEST:SCANner:MODE
:TEST:SCANner:TRIGger?
:TEST:SCANner:VALid?
Equivalent key
No equivalent key is available on the front panel.
sequence
298
:TEST:SCANner:MODE
:TEST:SCANner:MODE
on page 299
on page 300
on page 301
on page 302
on page 302
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
is set to ON, the interface pin signal value
should set to OFF so that scanner interface

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4981a/001E4981a/002

Table of Contents