Keysight E4981A Programming Manual page 93

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Measurement Applications (Sample Programs)
Executing OPEN/SHORT/LOAD Correction
' Measurement
'===================================
'===================================
' Setup End
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Corr_Result = FnComp("OPEN", 0.00002, Freq)
If Corr_Result <> 0 Then
MsgBox "OPEN compensation is not successful."
'End
End If
Corr_Result = FnComp("SHORT", 20, Freq)
If Corr_Result <> 0 Then
MsgBox "SHORT compensation is not successful."
'End
End If
Corr_Result = FnComp("LOAD", 0.2, Freq)
If Corr_Result <> 0 Then
MsgBox "LOAD compensation is not successful."
'End
End If
Corr_File_Name = "C:\E4981A_Corr_Data.txt"
' Save Data
Call Save_Corr_File(Corr_File_Name, Freq)
MsgBox "Correction data stored at " & Corr_File_Name