Command Reference
E4981A SCPI Commands

[:SENSe]:CORRection:OFFSet:DATA

Syntax
[:SENSe]:CORRection:OFFSet:DATA <numeric 1>,<numeric 2>
[:SENSe]:CORRection:OFFSet:DATA?
Description
Sets the correction values for the primary parameter and secondary parameter
used in the offset correction function.
These correction data are measured for the measurement frequency used in
executing the command (set with the
Parameters
Description
Range
Initial value
Unit
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
The unit of parameter changes depending on the limit range designation
method.
{numeric 1},{numeric 2}<newline><^END>
Query response
Related commands
[:SENSe]:CORRection:OFFSet[:STATe]
Equivalent key
[Meas Setup] - CORRECTION - OFFSET - A|B
sequence
246
<numeric 1>
The offset correction value for the
primary parameter.
-999.999 to 999.999
0
F (farad)
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
:SOURce:FREQuency[:CW]
<numeric 2>
The offset correction value for the
secondary parameter.
-99.9999E9 to 99.9999E9
0
Depends on the setup of the
secondary parameter.
on page 247
command).