Command Reference
E4981A SCPI Commands
:CALCulate1:COMParator:COUNt:MULTiple:OVLD?
Syntax
:CALCulate1:COMParator:COUNt:MULTiple:OVLD?
Description
Reads out the overload count value of the selected channel.
When the multi-correction function is turned off, reads out normal count value
(common value for all channels). In other words, this command has the same
function as the
only)
Query response
{numeric}<newline><^END>
Related commands
:CALCulate1:COMParator:COUNt:MULTiple:DATA?
[:SENSe]:CORRection:MULTiple[:STATe]
:CALCulate1:COMParator:COUNt[:STATe]
:CALCulate1:COMParator:COUNt:CLEar
:CALCulate1:COMParator[:STATe]
Equivalent key
No equivalent key is available on the front panel.
sequence
:CALCulate1:COMParator:COUNt:OVLD?
Syntax
:CALCulate1:COMParator:COUNt:OVLD?
Description
Reads out each overload count value. (Query only)
Query response
{numeric}<newline><^END>
Related commands
:CALCulate1:COMParator:COUNt:DATA?
:CALCulate1:COMParator:COUNt[:STATe]
:CALCulate1:COMParator:COUNt:CLEar
:CALCulate1:COMParator[:STATe]
Equivalent key
No equivalent key is available on the front panel.
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
:CALCulate1:COMParator:COUNt:OVLD?
command. (Query
on page 177
on page 245
on page 180
on page 175
on page 187
on page 176
on page 180
on page 175
on page 187
179