[:Sense]:Correction:data - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe]:CORRection:DATA

Syntax
[:SENSe]:CORRection:DATA {STAN1|STAN2|STAN3},<numeric 1>,<numeric 2>
[:SENSe]:CORRection:DATA? {STAN1|STAN2|STAN3}
Description
Sets the correction data for OPEN/SHORT/LOAD correction.
These correction data are measured for the measurement frequency used in
executing the command (set with the
When using the multi-correction function (set to ON with the
[:SENSe]:CORRection:MULTiple[:STATe]
correction data for the selected channel when executing the command
(selected with the
To activate the set correction data, use the
[:SENSe]:CORRection:OPEN[:STATe],
[:SENSe]:CORRection:SHORt[:STATe], and
[:SENSe]:CORRection:LOAD[:STATe]
OPEN/SHORT/LOAD correction function.
The transfer format of data read out with this command conforms to the setup
of the
Parameters
STAN1
STAN2
STAN3
Description
Range
Initial value
Unit
240
[:SENSe]:CORRection:MULTiple:CHANnel
command.
:FORMat[:DATA]
Parameter 1: {STAN1|STAN2|STAN3}
Sets or reads out the data for OPEN correction.
Sets or reads out the data for SHORT correction.
Sets or reads out the data for LOAD correction.
When specifying STAN1 as parameter 1:
Parameter 2: <numeric 1>
Value of the primary parameter
-99.9999E9 to 99.9999E9
0
Depends on the type of parameter.
1. Use the
[:SENSe]:CORRection:CKIT:STAN1:FORMat
specify the type of parameter.
:SOURce:FREQuency[:CW]
command), this is set as the
commands to turn on the
Parameter 3: <numeric 2>
Value of the secondary
1
.
parameter
-99.9999E9 to 99.9999E9
0
Depends on the type of parameter.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
command).
command).
*1
.
command to