Command Reference
E4981A SCPI Commands

:CALCulate1:MATH:STATe

Syntax
:CALCulate1:MATH:STATe {ON|OFF|1|0}
:CALCulate1:MATH:STATe?
Description
Determines whether to use the function (deviation measurement mode) that
displays the primary parameter measurement result in deviation from the
reference value (set using the
Changing any of the measurement parameters will automatically disable
this function. To avoid this, you have to write a program so that this
command is executed after the execution of the measurement parameter
setup command
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands
:CALCulate1:MATH:EXPRession:NAME
:CALCulate1:FORMat
:CALCulate2:FORMat
:CALCulate2:MATH:STATe
Equivalent key
[Meas Setup] - DEV A - OFF
sequence
190
:DATA[:DATA]
(:CALCulate1:FORMat
Description
Enables the deviation measurement mode.
Disables the deviation measurement mode (in other
words, displays the measurement result in an absolute
value).
on page 188
on page 191
on page 193
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
command).
and :CALCulate2:FORMat).
on page 189