Command Reference
E4981A SCPI Commands

[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2

Syntax
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2 <numeric>
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2?
Description
Sets/Gets the Threshold2 value for Contact check. Threshold2 is a parameter
which is proportional contact resistance of Lp or Lc. The recommended setting
value is described in the Operation Manual.
The contact failure at Lp can not be detected at the range of 220 uF to 1 mF
at 120 Hz and 22 uF to 100 uF at 1 kHz. At the same range, the threshold1
is used instead of this for Lc. (Therefore, the threshold2 is not used at the
range).
Parameters
Description
Range
Initial value
Resolution
Query response
<numeric><newline><^END>
Related commands
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe]
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1
Equivalent key
[Meas Setup] - CONT CHK- CC1 TH2
sequence
262
<numeric>
Threshold2 value for contact 1
0 to 1
1
0.01
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
on page 260
on page 261