9. Push the PC console back in as far as it will go. Lift up the front until
the side mounting holes are lined up with the threaded holes. Re-install
the 4 PC console mounting screws using the Phillips screwdriver. Also,
re-install the 4 screws that secure the PC to the rack.
10. Re-connect the 4 cables to the rear of the PC console.
11. Using the Phillips screwdriver, re-install the blank panel that was
previously removed above the PC console.
12. Slide the "Release" lever on the front of the PC console back to the
"Lock" position. Turn on the PC/Console circuit breaker on the
breaker box and re-start the PC to complete the procedure.
Use the following procedure to replace the National Instruments (NI)
Boards in the PC.
Equipment Required:
Phillips screwdriver
Boards
Flat blade screwdriver
NI Digital Board or NI Analog Board
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component. If an antistatic wrist
strap is not available, be sure to touch the instrument chassis before
Preventive Maintenance and Servicing

Replacing the National Instruments Boards

Omni FTIR Multi Gas CEMS Instruction Manual