Advanced Troubleshooting
Replacing the Covers of the TSQ 8000 Evo Instrument

Replacing the Chassis Feet

260
TSQ 8000 Evo Hardware Manual
To replace the chassis feet on the bottom of the TSQ 8000 Evo instrument
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
System"
on
page
107.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the TSQ 8000 Evo Instrument"
3. Follow the instructions in the section,
components required to safely replace the chassis feet.
4. Remove the small and large glass covers in the manifold.
5. Carefully tip the TSQ 8000 Evo instrument and rest it on its left side.
6. Use a T20 Torxhead screwdriver to remove the screws holding any damaged feet to the
bottom of the instrument.
Figure 277. Removing the Feet on the Bottom of the TSQ 8000 Evo Instrument
7. Attach the new chassis feet.
8. Set the instrument upright.
9. Replace the large and small glass manifold covers.
10. Follow the instructions in the section,
components you removed.
"Shutting Down the TSQ 8000 Evo
"Distribution Board"
on
Feet
"Distribution Board"
on
on
page
109.
page 300
to remove the
page 300
to replace the