Troubleshooting
Investigating Vacuum Issues
Turbomolecular pump shuts off while in use.
Table 23. Possible Remedies when Turbomolecular Pump Shuts Off While in Use
Run Leak Check diagnostics to look for leaks and address them. Excessive gas flow may cause the pump to overheat.
Ensure the temperature of the lab environment is 15-35 °C.
Make sure the combined flow of carrier gas and CI reagent gas to the TSQ 8000 Evo is less than 10 mL/min.