R&S AFQ100A

Baseband Block - Trigger and Input Settings

The Trigger and Input system contains the commands for configuration of the trigger settings.
Command
[SOURce:]INPut:CONTrol:IMPedance
[SOURce:]INPut:CONTrol:THReshold
[SOURce:]INPut:TRIGger:SEGMent:SLOPe
[SOURce:]INPut:TRIGger:SLOPe
[SOURce:]TRIGger:EXECute
[SOURce:]TRIGger:MODE
[SOURce:]TRIGger:RCOunt
[SOURce:]TRIGger:RENable
[SOURce:]TRIGger:RMODe?
[SOURce:]TRIGger:SEGment:DELay
[SOURce:]TRIGger:SEGMent:EXECute
[SOURce:]TRIGger:SEGMent:INHibit
[SOURce:]TRIGger:SEGMent:MODE
[SOURce:]TRIGger:SEGMent:RCOunt
[SOURce:]TRIGger:SEGMent:SOURce
[SOURce:]TRIGger:SOURce
[SOURce:]TRIGger[:EXTernal]:DELay
[SOURce:]TRIGger[:EXTernal]:INHibit
[SOURce:]WSEGment?
[SOURce:]WSEGment:NEXT
[SOURce:]WSEGment:NEXT:Mode
[SOURce:]INPut:CONTrol:IMPedance G50 | G1K
Clock source external only!
Sets the impedance for the three inputs CLOCK IN, TRIG, and NEXT (segment trigger). The lower
value is recommended for high clock frequencies. The meanings of the abbreviations G50 and G1k
are 0.05 ki and 1.00 ki, respectively.
Example:
INP:CONT:IMP G50
*RST value
G1K
1401.3084.32
Parameter
G50, G1K
0... 2 V
POSitive, NEGative
POSitive, NEGative
-
CONTinuous | SINGle | REPeatN
1...100
ON | OFF
0...10 sec
0...10 sec
CONTinuous | SEAMless | SINGle | REPeatN
1...100
MANual | EXTernal | BUS | AUTO
MANual | EXTernal | BUS | AUTO
0...10 sec
0...10 sec
-
-
AINCrement | FPARallel | FSERial
'sets the input impedance to 50 i.
Resolution
Options
-
AFQ Commands - Signal Generation
6.27
Default
Note
unit
-
V
-
-
No query
Query only
S
No query
S
S
S
Query only
SCPI
Device-specific
E-3