R&S AFQ100A Operating Manual page 348

I/q modulation generator
Hide thumbs Also See for AFQ100A:
Table of Contents

Advertisement

R&S AFQ100A
STATus:OPERation:EVENt?
The command queries the content of the EVENt part of the STATus:OPERation register. This part
contains information on the actions performed in the instrument since the last readout. The content
of the EVENt part is deleted after being read out.
The command is a query command and therefore has no *RST value.
Example:
STAT:OPER:EVEN?
*RST value
-
STATus:PRESet
The command resets the status registers. All PTRansition parts are set to FFFFh (32767), i.e. all
transitions from 0 to 1 are detected. All NTRansition parts are set to 0, i.e. a transition from 1 to 0 in
a CONDition bit is not detected. The ENABle parts of STATus:OPERation and
STATus:QUEStionable are set to 0, i.e. all events in these registers are not passed on.
The command triggers an event and therefore has no query form and no *RST value.
Example:
STAT:PRES
*RST value
-
STATus:QUEStionable:CONDition?
The command queries the content of the CONDition part of the STATus:QUEStionable register. This
part contains information on the action currently being performed in the instrument. The content is
not deleted after being read out since it indicates the current hardware status.
The command is a query command and therefore has no *RST value.
Example:
STAT:OPER:COND?
*RST value
-
1401.3084.32
'queries the STATus:OPERation:EVENt register.
Resolution
Options
-
'resets the status registers.
Resolution
Options
-
'queries the Status:Questionable:Condition register.
Resolution
Options
-