R&S AFQ100A
1401.3084.32
Run Mode "Repeat N Times"
The following segment starts to be output
immediately after a segment trigger event occurs.
After finishing the run of the last segment, signal
generation switches to idle mode. A segment
trigger event is without effect. Restart of signal
generation is triggered by a regular trigger event
activated automatically after a system-imposed
signal gap. The waveform is repeated as often as
the number of Repeat Count is set (see
Count -
last run signal generation stops.
Remote-control command:
SOUR:TRIG:SEGM:MODE CONT
The current segment is output continuously (i.e.
Continuous
cyclically) until the next segment trigger event
Seamless
occurs.
Signal output will switch to the next segment
when the current segment is output completely to
achieve wrap around, seamless signal transition.
Run Mode "Continuous"
When a segment trigger event occurs, the
following segment is output after finishing the
current one.
When the run of the last segment is finished,
signal generation starts immediately at the
beginning, provided that a segment trigger event
has been occured before. Otherwise signal
output switches to idle mode and starts again
with the next trigger or segement trigger event.
4.19
Baseband Block
Trigger, on page 4.15). After finishing the
Repeat
E-3