Divider - Marker
Pattern - Marker
1401.3084.32
A marker signal that is defined by a pattern of
List
samples is generated. When List is selected an
input box for entering the pattern is displayed (see
also
List
The samples can be assorted by entering either
particular samples or sample ranges, each defined
by the sampleindex and the signal value (0/1). The
arguments are separated by:
A hyphen - between the start and the end
A colon : between the sample index and the
A semicolon ; between the argument entries.
Example:
Remote-control command:
SOUR:TRIG:OUTP1:MODE LIST
SOUR:TRIG:OUTP1:LIST
"1-100:1;101:0;102-200:1"
(Marker signal output Pulse only)
Enters the divider value.
The precision of the frequency setting depends on the sampling rate.
The maximum pulse frequency is equal to half of the sampling rate.
The entry field is indicated only if a regular marker signal (Pulse) is
generated. The resulting pulse frequency is displayed below.
Remote-control command:
SOUR:TRIG:OUTP1:MODE PULS
SOUR:TRIG:OUTP1:PULS:DIV 4
SOUR:TRIG:OUTP1:PULS:FREQ?
(Marker signal output Pattern only)
Set a pattern for the marker output signal. The pattern has a maximum
length of 32 bits.
The pattern bits switch the marker signal to high and low state.
Remote-control command:
SOUR:TRIG:OUTP1:MODE PATT
SOUR:TRIG:OUTP1:PATT #11F