R&S AFQ100A
CALibration:AOUTput<CH>:DELay?
The command adjusts the amplitude of the (selected) analog output.
The instrument generates a series of signal levels and converts them using a precise ADC. The
procedure takes about 30 seconds.
The command is a query command and therefore has no *RST value.
Example:
CAL:AOUT:AMPL?
Response: '0'
*RST value
-

Diagnostic Subsystem

The DIAGnostic system contains the commands used for instrument diagnosis and servicing.
In the Test Point section provides access to the test points available in the instrument except for the
analog output board (see below). When activated, the voltage of the selected test point is measured
and displayed. A detailed description of the test points can be found in chapter 3 of the service manual.
In the Analog Output Board Test Board section provides access to the test points on the analog output
board
In the Fan Control section the fan speed is selected.
Command
DIAGnostic:AOUTput<CH>:POINt:CATalog?
DIAGnostic:AOUTput<CH>:POINt?
DIAGnostic:BGINfo?
DIAGnostic:FCONtrol
DIAGnostic:INFO:POCounter?
DIAGnostic:INFO:OTIMe?
DIAGnostic[:MEASure]:POINt?
DIAGnostic:POINt:CATalog?
1401.3084.32
'starts all internal adjustments for which no external measuring equipment is
needed. The adjustments with external measuring equipment are described in
the service manual.
'calibration was successful
Resolution
Options
-
AFQ Commands - Setup Section
Parameters
<module name>
AUTO, FSPeed
6.73
SCPI
Device-specific
Remark
Default
unit
Query only
Query only
Query only
Query only
Query only
Query only
Query only