R&S AFQ100A Operating Manual page 312

I/q modulation generator
Hide thumbs Also See for AFQ100A:
Table of Contents

Advertisement

R&S AFQ100A
[SOURce:]EQUalizer:IQ:Q:FREQuency 0 ... 1 GHz
This command enters frequency values into the current data table of the Q path.
The command triggers an event and therefore has no query form and no *RST value.
Example:
EQU:IQ:Q:FREQ 10000,20000,30000
*RST value
-
[SOURce:]EQUalizer:IQ:Q:GAIN -5 ... +50 dB
This command enters gain values into the current data table of the Q path.
The command triggers an event and therefore has no query form and no *RST value.
Example:
EQU:IQ:Q:GAIN 0,-5,-3
*RST value
-
[SOURce:]EQUalizer:IQ:Q:PHASe -360 ... +360 deg
This command enters phase values into the current data table of the Q path.
The command triggers an event and therefore has no query form and no *RST value.
Example:
EQU:IQ:Q:PHAS 0,-5,-10
*RST value
-
[SOURce:]EQUalizer:IQ:Q:POINts?
This command queries the number of data points of the current data table of the I path.
The command is a query command and therefore has no *RST value.
Example:
EQU:IQ:I:SEL "D:/TEMP/TESTPREDIQ"
EQU:IQ:Q:POIN?
*RST value
-
1401.3084.32
'enters three frequency values into the current data table.
Resolution
Options
1 Hz
'enters three gain values into the current data table.
Resolution
Options
0.001 dB
'enters three gain values into the current data table.
Resolution
Options
0.01 deg
'selects the data table.
'queries the number of data points of the current data table.
Resolution
Options
-
AFQ Commands - Signal Generation
6.51
SCPI
Device-specific
SCPI
Device-specific
SCPI
Device-specific
SCPI
Device-specific
E-3