R&S AFQ100A Operating Manual page 290

I/q modulation generator
Hide thumbs Also See for AFQ100A:
Table of Contents

Advertisement

R&S AFQ100A
[SOURce:]TRIGger:MODE CONTinuous | SINGle | REPeatN
The command selects the trigger mode. i.e. continuous, single, or repeating the waveform for a
defined number of outputs. After a trigger event, the waveform will be output continuously or once.
Continuous
Parameter:
Single
Repeat N Times
Example:
TRIG:MODE SING
*RST value
CONTinuous
1401.3084.32
The waveform is output continuously.
Retrigger "Off"
When retrigger is disabled a subsequent trigger after the start of signal
generation is without effect:
The waveform output can be restarted by switching off and then switching on
the ARB state in the Baseband Block.
Retrigger "On"
When retrigger is enabled, the current signal output is reset at a subsequent
trigger after start of signal generation.
The waveform is output once after a trigger event occured. Every subsequent
trigger event (internal by clicking the Execute button or external) causes a
restart.
Retrigger "Off"
When retrigger is disabled a trigger during signal generation is without effect:
Retrigger "On"
When retrigger is enabled the current signal output is reset at a trigger during
signal generation.
Multiple output of the waveform when a trigger event occurs.
Note:
Run mode Repeat N Times is enabled only in combination with Segment
Trigger Source Auto. If Segment Trigger Source is set different to Auto,
the instrument operates in run mode Continuous.
Retrigger "Off"
When retrigger is disabled a trigger during signal generation is without effect:
Retrigger "On"
When retrigger is enabled the current signal output is reset at a subsequent
trigger event after start of signal generation.
The waveform is repeated as often as the number of Repeat Count is set.
After finishing the last run signal generation stops.
The waveform output can be restarted by retriggering or, switching off and
then switching on the ARB state in the baseband block.
'the waveform is output once only when a trigger event occurs.
Resolution
Options
-
AFQ Commands - Signal Generation
6.29
SCPI
Device-specific
E-3