Self Test Failed; Settings Conflict; String Data Not Allowed (-158); Suffix Not Allowed (-138) - R&S AFQ100A Operating Manual

I/q modulation generator
Hide thumbs Also See for AFQ100A:
Table of Contents

Advertisement

R&S AFQ100A

Self test failed ... (-330)

An error was detected in the selftest named after the semicolon. An error-free operation of the module
concerned is no longer guaranteed.
Device specific error - sets bit 3 in the ESR register.
SCPI:

Settings conflict ... (-221)

There is a setting conflict between the two parameters indicated after the semicolon.
The set FSK deviation is too large for the selected symbol rate.
Example:
One of the given values has to be corrected to obtain a valid output signal.
Remedy:
Execution error - sets bit 4 in the ESR register.
SCPI:

String data not allowed (-158)

The command contains a legal string data element which is not allowed at this point.
A text parameter is set in quotation marks - SOURce:FREQuency:MODE
Example:
"FIXed".
Command error - sets bit 5 in the ESR register.
SCPI:

Suffix not allowed (-138)

A suffix is not allowed for this command or at this point of the command.
Command *RCL does not permit indicating a suffix.
Example:
Command error - sets bit 5 in the ESR register.
SCPI:

Suffix too long (-134)

The suffix contains more than 12 characters.
Command error - sets bit 5 in the ESR register.
SCPI:
1401.3084.32
Alphabetical List of SCPI-Error Messages
9.8
E-3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Afq100b

Table of Contents