In the 'Measurement & Automation Control' program, this setting is made on the Template tab.
VXI-11 Interface Messages
On the Ethernet link, the interface messages are called low-level control messages. These messages
can be used to emulate interface messages of the IEC/IEEE bus.
Command
&ABO
(Abort)
&DCL
(Device Clear)
&GTL
(Go to Local)
&GTR
(Go to Remote)
&GET
(Group Execute Trigger)
&LLO
(Local Lockout)
&POL
(Serial Poll)
&NREN
(Not Remote Enable)
Table 5-9
VXI-11 Interface Messages
1401.3084.32
Remote Control Hardware Interfaces
Effect on the instrument
Aborts the processing of the commands just received.
Aborts the processing of the commands just received and sets the command
processing software to a defined initial state. Does not change the instrument
setting.
Transition to the "Local" state (manual control)
Transition to the "Remote" state (remote control)
Triggers a previously active device function (eg a sweep). The effect of the
command is the same as with that of a pulse at the external trigger signal
input.
Disables switchover from remote control to manual control by means of the
front panel keys
Starts a serial poll
Enables switchover from remote control to manual control by means of the
front panel keys