[SOURce:]INPut:BERT:THReshold 0 ... 2.0 V
The command sets the high/low threshold of the BERT inputs Clock, Data, Data Enable and
Restart. In the case of positive polarity, this threshold determines the point as of which a signal is
high (active) or low (inactive).
Example:
INP:BERT:THR 1 V
*RST value
1.00 V
TRIGger:BERT:IMMediate
The command triggers a BERT measurement.
This command triggers an event and therefore has no *RST value and no query form.
Example:
TRIG:BERT:IMM
*RST value
-
1401.3084.32
'a high/low threshold of 1 volt is set at the BERT inputs. The signal is high
(active) for a signal voltage of 1 volt and higher.