R&S AFQ100A
Analog output calibration only applies to the attenuator mode.
The instrument generates a series of signal levels and converts them using a precise ADC. The
procedure takes about 30 seconds.
All
Adjust Amplitude
Adjust Delay
1401.3084.32
Starts all internal adjustments for which no external measuring
equipment is needed.
Note:
Adjustments with external measuring equipment are described in
chapter 2 of the service manual, section Adjustments.
Remote-control command:
CAL:ALL?
Starts every internal adjustment procedures which affect the amplitude
at the analog output.
Remote-control command:
CAL:AOUT:AMPL?
Starts every internal adjustment procedures which affect the delay at
the analog output.
Remote-control command:
CAL:AOUT:DEL?
4.84
General Instrument Settings
E-3