[SOURce:]TRIGger:OUTPut<[1]...4>:DELay:MAXimum?
The command queries the maximum marker delay for setting :TRIGger:OUTPut:DELay:FIXed
ON.. The command is a query only and therefore has no *RST value.
Example:
TRIG:OUTP:DEL:FIX ON
TRIG:OUTP:DEL:MAX?
Response: "4000"
*RST value
-
[SOURce:]TRIGger:OUTPut<[1]...4>:DELay:MINimum?
The command queries the minimum marker delay for setting :TRIGger:OUTPut:DELay:FIXed
ON..
The command is a query only and therefore has no *RST value.
Example:
TRIG:OUTP:DEL:FIX ON
TRIG:OUTP:DEL:MIN?
Response: "0"
*RST value
-
[SOURce:]TRIGger:OUTPut<[1]...4>:LIST
The command defines the pattern of samples used to generate the marker signal in the setting
TRIGger:OUTPut:MODE LIST for the selected marker output.
Example:
SOUR:TRIG:OUTP1:LIST "1-100:1;101:0;102-200:1"
Alias remote-control command:
SOUR:MARK1:LIST "1-100:1;101:0;102-200:1"
*RST value
UNCHanged
1401.3084.32
'restricts the marker signal delay setting range to the dynamic range.
'queries the maximum of the dynamic range.
'the maximum for the marker delay setting is 4000 samples.
Resolution
Options
-
'restricts the marker signal delay setting range to the dynamic range.
'queries the minimum of the dynamic range.
'the minimum for the marker delay setting is 0 samples.
Resolution
Options
-
'The signal starts with value 1 from position 1 (start index) to position 100 (end
index). By the sample position 101 the signal changes to 0 and during the
sample range positions 102 to 200 the signal is set to 1 again.
Resolution
Options
-