[SOURce:]EQUalizer:IQ:CATalog?
This command lists all files containing I/Q equalizer data tables saved on local hard disk. This
command reads out the subdirectories and files in the specified directory. The default directory is set
using command MMEM:CDIRectory.
The command is a query command and therefore has no *RST value.
Example:
MMEM:CDIR 'd:/user/Equalizer'
EQU:IQ:CAT?
*RST value
-
[SOURce:]EQUalizer:IQ:DELete <FILE NAME>
This command deletes the selected I/Q equalizer data table from local hard disk.
The command triggers an event and therefore has no query form and no *RST value.
Example:
*RST value
-
[SOURce:]EQUalizer:IQ:I:FREQuency 0 ... 1 GHz
This command enters frequency values into the current data table of the I path selected using the
EQU:IQ:I:SEL command.
Example:
EQU:IQ:I:FREQ 10000,20000,30000
*RST value
-
[SOURce:]EQUalizer:IQ:I:GAIN -5 ... +50 dB
This command enters gain values into the current data table of the I path.
Example:
EQU:IQ:I:GAIN 0,-5,-3
*RST value
-
1401.3084.32
'sets the default directory
'lists all files containing I/Q equalizer data tables.
Resolution
Options
-
EQU:IQ:SEL "iqdata_1"
'selects the file 'iqdata_1.afq_iq_fgp' to be deleted.
EQU:IQ:DEL "iqdata_1"
'deletes the file 'iqdata_1.afq_iq_fgp' from HD.
Resolution
Options
-
'writes three frequency values into the current data table.
Resolution
Options
1 Hz
'writes three gain values into the current data table.
Resolution
Options
0.001 dB