[SOURce:]ROSCillator:SYNC:MODE NORMal | MASTer | SLAVe | LAST
The command selects the mode for synchronization with other R&S AFQ waveform generators.
Note:
The synchronization of several connected AFQ waveform generators requires the baseboard
version 1.6 or higher. Up to baseboard version 1.5 only synchronization mode Normal is
provided (stand-alone mode).
NORMal
Parameters:
MASTer
SLAVe
LAST
Example:
ROSC:SOUR:SYNC:MODE SLAV
*RST value
NORMal
1401.3084.32
The instrument is working in stand-alone mode.
The instrument provides all connected instruments with its reference clock
signal.
The instrument receives the clock signal from another AFQ waveform
generator (Master).
This instrument receives the clock signal from another AFQ waveform
generator (Master). It is appointed to be the last one in the chain of the
connected AFQs.
'sets the instrument to "Slave". The instrument receives the reference
oscillator signal from another AFQ waveform generator (Master) for
synchronization.
Resolution
Options
-
AFQ Commands - Signal Generation
6.44
SCPI
Device-specific
E-3