R&S AFQ100A
[SOURce:]TSIGnal:SINE:CREate:NAMed <waveform filename>
The command generates a signal and stores the file on the local hard disk (HD).
The Create Test Signal window opens automatically and the signal can be stored as a waveform
file.
This command triggers an event and therefore has no *RST value and no query form.
Example:
TSIG:SINE:CRE:NAM "D:\temp\test.wv"
*RST value
-
[SOURce:]TSIGnal:SINE:FREQuency?
The command queries the resulting frequency of the test signal.
The command is a query command and therefore has no *RST value.
Example:
TSIG:SINE:FREQ?
Response: "80000"
*RST value
-
[SOURce:]TSIGnal:SINE:PHASe -180 ... 180 degree
The command sets the phase offset of the sinewave signal on the Q channel relative to the
sinewave signal on the I channel.
Example:
TSIG:SINE:PHAS 45
*RST value
-90
[SOURce:]TSIGnal:SINE:SAMPles 4 ... 1000 samples
The command sets the number of sample values required from the sine wave per period.
The resulting clock rate must not exceed the maximum ARB clock rate of 300 MHz.
The number of sample values is automatically restricted by reference to the set frequency.
Example:
TSIG:SINE:SAMP 500
*RST value
100
1401.3084.32
'writes the generated waveform file to 'test.wv' on the HD.
Resolution
Options
'queries the resulting frequency of the test signal.
'the frequency amounts 80000 kHz.
Resolution
Options
'sets the phase offset of the Q channel to 45 degree.
Resolution
Options
'sets the number of samples to 500 for a period.
Resolution
Options
AFQ Commands - Signal Generation
6.13
SCPI
Device-specific
SCPI
Device-specific
SCPI
Device-specific
SCPI
Device-specific
E-3