R&S AFQ100A
"DIAG:BGIN:CAT?"
Example:
Response: MBRD,SATT3C,BBINS,BBINR,BBOUT,....
"DIAG:BGIN? 'MBRD'"
Response:
DIAGnostic:FCONtrol AUTO | FSPeed
The command sets the fan speed of the instrument. The full speed mode is provided for service
purposes.
Example:
DIAG:FCON FSP
*RST value
AUTO
DIAGnostic:INFO:OTIMe?
The command queries the number of operation hours
"DIAG:INFO:OTIM?"
Example:
Response:
*RST value
Resolution
-
DIAGnostic:INFO:POCounter?
The command queries the number of power-on events.
"DIAG:INFO:POC?"
Example:
Response:
*RST value
Resolution
-
1401.3084.32
'queries the instrument configuration of path A..
'returns the data of all available modules.
'queries the configuration of the motherboard of path A.
MBRD 1141.3501.02 1.5.3 100023
'module motherboard with stock number 1141.3501.01 has revision 1.5.3 and
serial number 100023.
'sets full fan speed.
Resolution
Options
-
'queries the operation hours.
"100023"
'the instrument was operated for 100023 hours up to now.
Options
-
-
'queries the number of power on events.
"123"
'the instrument was switched on for 123 times up to now.
Options
-
-