R&S AFQ100A
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess 1 ... 30
The command sets the IEC/IEEE-bus address.
Example:
SYST:COMM:GPIB:ADDR 14
*RST value
28
SYSTem:COMMunicate:USB:RESource?
The command queries the visa resource string for remote control via the USB interface. .
The command is a query command and therefore has no *RST value.
Example:
SYST:COMM:USB:RES?
Response:
*RST value
SYSTem:DATE <year>,<month>,<day>
The command sets the date for the instrument-internal calendar.
Example:
SYST:DATE 2005,11,01
*RST value
-
SYSTem:DISPlay:UPDate ON | OFF
The command switches the update of the display on/off. A switchover from remote control to manual
control always sets the status of the update of the display to ON.
Example:
SYST:DISP:UPD OFF
*RST value
ON
1401.3084.32
'sets IEC/IEEE-bus address 14.
Resolution
Options
-
'queries the VISA resource string for remote control via the USB interface.
USB::72::000000::INSTR
Resolution
Options
-
'sets November 1, 2005.
Resolution
Options
-
'switches update of displayed parameter values off.
Resolution
Options
-
AFQ Commands - Setup Section