R&S AFQ100A Operating Manual page 125

I/q modulation generator
Hide thumbs Also See for AFQ100A:
Table of Contents

Advertisement

R&S AFQ100A
1401.3084.32
When retrigger is enabled the current signal output
is reset at a trigger during signal generation.
Remote-control command:
SOUR:TRIG:MODE SING
Multiple output of the waveform when a trigger event
Repeat N
occurs.
Times
Note:
Run mode Repeat N Times is enabled only in
combination with Segment Trigger source Auto
(see
If Segment Trigger source is set different to
Auto, the instrument operates in run mode
Continuous.
Retrigger "Off"
When retrigger is disabled a trigger during signal
generation is without effect:
Retrigger "On"
When retrigger is enabled the current signal output
is reset at a subsequent trigger event after start of
signal generation.
The waveform is repeated as often as the number of
Repeat Count is set (see
on page 4.34). After finishing the last run signal
generation stops.
The waveform output can be restarted by
retriggering or, switching off and then switching on
the ARB state in the baseband block.
Remote-control command:
SOUR:TRIG:MODE REP
4.14
"Source - Segment
Trigger", on page 4.22).
Repeat Count -
Baseband Block
Trigger,
E-3