Program Mnemonic Too Long (-112); Queue Overflow (-350); Query Interrupted (-410); Query Unterminated (-420) - R&S AFQ100A Operating Manual

I/q modulation generator
Hide thumbs Also See for AFQ100A:
Table of Contents

Advertisement

R&S AFQ100A

Program mnemonic too long (-112)

The header contains more than 12 characters.
Command error - sets bit 5 in the ESR register.
SCPI:

Queue overflow (-350)

This error code is entered into the queue instead of the actual error code if the queue is full. It indicates
that an error has occurred but not been recorded in the queue. The original error message is lost.
Reading out the error messages e.g. with command SYSTem:ERRor:ALL?
Remedy:
clears the error queue.
Device specific error- sets bit 3 in the ESR register.
SCPI:

Query interrupted (-410)

This query has been interrupted.
After a query, the instrument receives new data before the response has been
Example:
sent completely.
Query error - error in data request - sets bit 2 in the ESR register.
SCPI:

Query unterminated (-420)

This query is missing or incomplete.
The instrument is addressed as a talker and receives incomplete data.
Example:
Query error - error in data request - sets bit 2 in the ESR register.
SCPI:

Query deadlocked (-430)

This query cannot be processed.
The input and output buffers are full, the instrument cannot be operated.
Example
Query error - error in data request - sets bit 2 in the ESR register.
SCPI:
1401.3084.32
Alphabetical List of SCPI-Error Messages
9.7
E-3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Afq100b

Table of Contents