R&S AFQ100A Operating Manual page 295

I/q modulation generator
Hide thumbs Also See for AFQ100A:
Table of Contents

Advertisement

R&S AFQ100A
[SOURce:]TRIGger[:EXTernal]:DELay 0 ... 10 sec
The command specifies the trigger delay for external triggering. It enables the instrument to be
synchronized with the device under test or other external devices. A delay can be entered only when
an external trigger source has been selected.
Example:
TRIG:SOUR EXT
TRIG:DEL 10
*RST value
0
[SOURce:]TRIGger[:EXTernal]:INHibit 0 ... 10 sec
The command specifies the time interval by which a restart is to be inhibited following a trigger
event. This command applies only in the case of external triggering.
Example:
TRIG:SOUR EXT
TRIG:INH 2
*RST value
0
[SOURce:]WSEGment?
This query command indicates which segment the instrument is generating.
The command is a query command and therefore has no *RST value.
Example:
SOUR:WSEG?
*RST value
-
1401.3084.32
'selects an external trigger via the TRIG connector.
'sets a trigger delay of 10 seconds.
Resolution
Options
10 ns
'selects an external trigger via the TRIG connector.
'sets a restart inhibit for 2 seconds following a trigger event.
Resolution
Options
10 ns
'indicates which segment the instrument is generating.
Resolution
Options
-
AFQ Commands - Signal Generation
6.34
SCPI
Device-specific
SCPI
Device-specific
SCPI
Device-specific
E-3

Hide quick links: