Enable Remote Desktop Connection on R&S AFQ
The configuration is performed in the Windows XP menus. The operating system can only be accessed
if an external keyboard and monitor is connected. A mouse is recommended for convenient operation of
Windows XP. To ensure recognition of the external devices, the instrument has to be switched off prior
to connecting them.
1.
Switch off the instrument.
2.
Connect the external keyboard, mouse and monitor to the USB interface.
3.
Switch on the instrument.
4.
Open the start menu.
5.
Select My Computer and open the context menu by pressing the right mouse key.
6.
Click on Properties and select the Remote tab in the menu.
7.
Activate check box Allow users to connect remotely to this computer and close window with the
OK button.
Note:
When activating/locking the Remote Desktop, the associated firewall settings are automatically
enabled/disabled.