Help Menu - R&S AFQ100A Operating Manual

I/q modulation generator
Hide thumbs Also See for AFQ100A:
Table of Contents

Advertisement

R&S AFQ100A
Note:
Since Factory Preset resets the remote channel and network settings to the default values,
executing Factory Preset via remote control terminates the connection to the instrument, if these
settings had been configured to values different to the default ones!
The Factory Preset function resets nearly all instrument settings. In addition to the regular preset by
means of the
key, a Factory Preset resets also the following values:
PRESET
Reference frequency settings (Ref Oscillator dialog)
Network settings including hostname (Setup dialog)
Remote Channel settings including IEC/IEEE bus address and emulation (Setup dialog)
Start/Stop Gui Update (Setup dialog)
Display and keyboard settings (Setup dialog).
To maintain security, password settings and all settings protected by these passwords like disabled USB
and LAN connections are not changed.
Not affected by the Factory Preset are also user data, lists or instrument settings files, created for
example by means of the Save/Recall function.
Remote-control command:
SYST:FPR

Help Menu

The help system offers comprehensive online help for the instrument. A desired topic can be selected
via the table of contents (select Manual) or the index (select Index).
For context-sensitive information about a marked parameter, press the
1401.3084.32
General Instrument Settings
F1
4.101
key.
E-3